Bilgilendirme: Kurulum ve veri kapsamındaki çalışmalar devam etmektedir. Göstereceğiniz anlayış için teşekkür ederiz.
 

Designing and In Vitro Testing of a Novel Patient-Specific Total Knee Prosthesis Using the Probabilistic Approach

Loading...
Publication Logo

Date

2022

Authors

Korkmaz, Ismail H.
Kaymaz, Irfan
Yildirim, Omer S.
Murat, Fahri
Kovaci, Halim

Journal Title

Journal ISSN

Volume Title

Publisher

Walter de Gruyter GmbH

Open Access Color

OpenAIRE Downloads

OpenAIRE Views

Research Projects

Journal Issue

Abstract

In order to prevent failure as well as ensure comfort, patient-specific modelling for prostheses has been gaining interest. However, deterministic analyses have been widely used in the design process without considering any variation/uncertainties related to the design parameters of such prostheses. Therefore, this study aims to compare the performance of patient-specific anatomic Total Knee Arthroplasty (TKA) with off-the-shelf TKA. In the patient-specific model, the femoral condyle curves were considered in the femoral component's inner and outer surface design. The tibial component was designed to completely cover the tibia cutting surface. In vitro experiments were conducted to compare these two models in terms of loosening of the components. A probabilistic approach based on the finite element method was also used to compute the probability of failure of both models. According to the deterministic analysis results, 103.10 and 21.67 MPa von Mises stress values were obtained for the femoral component and cement in the anatomical model, while these values were 175.86 and 25.76 MPa, respectively, for the conventional model. In order to predict loosening damage due to local osteolysis or stress shield, it was determined that the deformation values in the examined cement structures were 15% lower in the anatomical model. According to probabilistic analysis results, it was observed that the probability of encountering an extreme value for the anatomical model is far less than that of the conventional model. This indicates that the anatomical model is safer than the conventional model, considering the failure scenarios in this study.

Description

Korkmaz, Ismail Hakkı/0000-0003-2440-0319; Kovacı, Halim/0000-0002-9053-3593;

Keywords

Experimental Validation, Finite Element Analysis, Probability-Based Design, Total Knee Arthroplasty

Fields of Science

Citation

WoS Q

Q3

Scopus Q

Q3

Source

Biomedical Engineering-Biomedizinische Technik

Volume

67

Issue

4

Start Page

295

End Page

305
Google Scholar Logo
Google Scholar™

Sustainable Development Goals

SDG data could not be loaded because of an error. Please refresh the page or try again later.